A. Yes it is possible to differentiate between ethylene glycol and propylene glycol using both refractive index and specific gravity.
You will need:
Digital refractometer for glycol, hydrometer 1.000-1.500 with subdivisons 0.005, hydrometer jar 300x50mm
[caption id="attachment_12965" align="alignleft" width="321"] Digital refractometer for glycol[/caption]
The key to differentiate the two is to use the physical differences with regard to specific gravity and refractive index.
Specific Gravity
Perform the specific gravity test first. Fill the cylinder with 250-300ml of the unknown sample. Record the numerical value of the nearest graduation mark, if the value is 1.05 or above the solution is ethylene glycol based. It is impossible for propylene glycol solutions to have a specific gravity at or above 1.045 at 70 degrees fahrenheit.
Refractive Index
If the Specific Gravity is between 1.0 and 1.05 (0.05 margin of error added), then you must do the Refractive Index to determine which glycol is present. We recommend using a hand held refractometer such as the one shown above. To perform this test all you have to do is place a few drops of your unknown sample liquid on the measurement prism at the end of the refractometer. Close the cover so the liquid spreads evenly across the entire surface of the prism without any air bubbles or dry spots. Allow this to remain on the prism for approximately 30 seconds.
It is best to use the instructions that come with the refractometer. Using a hypothetical Specific Gravity of 1.03 from the test above. A Specific Gravity of 1.03 means that you either have an Ethylene Glycol solution between 25% and 30% by weight, or you have a Propylene Glycol solution between 35% and 40% by weight. Which is it? In the refractometer field of view below you can see an area of grey scale and an area of white. The junction of these two determines the percent strength of a known Glycol. Here it shows that the solution could be 30% Propylene Glycol. Or perhaps it could be approximately 30% Ethylene Glycol.
Use the refractometer to identify an unknown glycol which we know to have two totally different percentages of strength. In our hypothetical example, if the grey and white shaded areas meet near the red line (red line added for the purpose of demonstration) our unknown is a Propylene Glycol solution. If, however, the grey and white shaded areas meet near the blue line (blue line added for the purpose of demonstration) our unknown is an Ethylene Glycol solution.
Thanks for sharing such a great information.. It really helpful to me.. I always search to read the quality content and finally i found this in you post. keep it up!
I came across your article about "DPD Tablets" on Camlab Info, and it's a valuable resource for anyone seeking information about water testing and analysis using DPD tablets. Your explanation of their usage and significance in water quality monitoring is quite informative.
This article on nanocolor tube tests is a gem for analytical enthusiasts. The idea of zero calibration is intriguing, and your detailed explanation makes it accessible. The step-by-step process helps demystify the technique for beginners like me. The use cases you've highlighted, particularly in...
Grant have introduced the new Controlled Rate Freeze Thaw (CRFT), which is a liquid nitrogen and cryogen free controlled rate freeze and thaw system for cryopreservation in a wide range of applications. Cryopreservation is a process where biological materials such as cells, tissues...
A dipslide is a means of testing the microbial content of liquids. This is a common method of monitoring and measuring microbial activity within many systems and commonly cooling systems. Usually the use of a dipslide is the first line of defence for the testing of microbial issues that could be...
Purchasing items using donated funds without the burden of VAT will now save many Millions of pounds each year which can be put to better use in advancing scientific knowledge to provide new treatments and cures that will shape the medicines of the future.
Working with pressure/vacuum conditions in the lab is common but not all products are designed to withstand this. Products with appropriate geometry and wall thickness are suitable for working safely with pressure. For example, filtering flasks, desiccators, or flat flange vessels, as well as...