How to connect the anti static set for Burkle pumps

Loading... 4 view(s)
How to connect the anti static set for Burkle pumps
The anti-static set for Burkle pumps is essential when decanting flammable liquids. Without a properly earthed system static charge can build up, which can result in fire or even explosion. Connect the Burkle anti static set correctly following the instructions below [caption id="attachment_10283" align="aligncenter" width="419"] The Burkle Anti-static Set for use with their liquid pumps[/caption]

Setup 1 - decanting into a conductive container

If you are pumping flammable liquids into a conductive container, use the black wire to connect the bulk container to earth (e.g. via a water pipe) and use the red wire to connect the two containers together. Step 1

Setup 2 - decanting into a non-conductive container

If the receiving container is not conductive, use the black wire to connect the bulk container to earth (e.g. via a water pipe) and use the bare copper wire to connect the bulk container to the decanted liquid. Step 2   For more information on the anti static set and the related Burkle liquid pumps, follow the links above or contact us if you have any questions about safe liquid transfer;