TitroLine 7000 automatic titrator for greater control and capacity for high accuracy work.
Used in a range of industries from analytical laboratory work to food, water, wastewater or environmental testing.
Automatic Titrator for exchangeable heads for potentiometry
With its performance spectrum, the TitroLine 7000 is the ideal starting device for potentiometric titration with potential for expansion and automation. Thanks to the high-resolution and precise pH/ mV and “dead-stop” measuring interface, it is possible to determine a wide range of parameters quickly, reliable and accurate.
General features
Specific features for the TitroLine 7000
Latest features
Measuring input analog | pH/mV-input with 24 bit transducer. Electrode socket according to DIN 19 262 or BNC with RFID receiver for SI Analytics ID electrodes |
Measuring range pH/accuracy | - 3.0… 18.00 / 0.002 +/- 1 digit |
Measuring range mV/accuracy | - 2000…2000/ 0.1 +/- 1 digit |
Measuring input µA: | μA-connector for double platinum electrodes, connection: 2 x 4 mm - sockets), polarisation voltage variably adjustable |
Measuring input temperature: | Connector for Pt 1000 / NTC 30 KOhm ( connection: 2 x 4 mm - sockets) |
Measuring range temperature | Pt 1000 - 75…195 °C / NTC 30 KOhm - 40 …125 °C |
Interfaces: | 3 × USB (2 x USB-A and 1 × USB-B), 2 × RS-232-C and 1 x LAN |
Interchangeable burettes | 5, 10, 20 ml or 50 ml intelligent interchangeable burette units |
Resolution (Steps): | 20.000 steps |
Dosing accuracy: | according to DIN ISO 8655, part 3: accuracy: < 0.15 % precision: < 0.05 - 0.07 % (in dependence of the used burette size) |
Display | 3.5 inches - 1/4 VGA TFT graphic color display with 320 x 240 pixels |
Housing material | Polypropylene |
Dimensions | 15.3 x 45 x 29.6 cm (W x H x D), height with interchangeable head |
Front keyboard | Polyester coated |
Material interchangeable burettes | Valve: PTFE/ETFE; Cylinder: borosilikate glass 3.3 (DURAN®) hoses (FEP, blue) |
Weight | 2.3 kg for basic device, 3.5 kg für complete device with interchangeable head (with empty reagent bottle) |
Ambient conditions | Ambient temperature: + 10 ... + 40 °C for operation and storage Humidity according to EN 61 010, Part 1: max. relative humidity 80 % for temperatures up to 31 °C, linear decrease down to 50 % relative humidity at a temperature of 40 °C |
Power supply: | 100-240 V; 50/60 Hz, power input: 30 VA |
CE-sign: | EMC compatibility according to the Council Directive: 2004/108/EG; applied harmonized standards: EN 61326-1:2006 Low-voltage directive according to the Council Directive 2006/95/EG Testing basis EN 61 010, Part 1 |