£360.78 excl. VAT
£432.94 inc. VAT
The MD 100 photometer is a convenient choice for testing up to 14 parameters on a single instrument.
The instrument has a simple user interface and is constructed for field use. Several industry specific versions are available the combine all of the tests you need.
Optics | LEDs, interference filters (IF) and photo sensor in transparent sample chamber. Depending on the version, up to 3 different interference filters are used. Wavelength specifications of interference filters: 430 nm Δλ = 5 nm 530 nm Δλ = 5 nm 560 nm Δλ = 5 nm 580 nm Δλ = 5 nm 610 nm Δλ = 6 nm 660 nm Δλ = 5 nm |
Wavelength Accuracy | ± 1 |
Photometric Range | -2600 - 2600 mAbs |
Photometric Accuracy | 3 % FS (T = 20 °C – 25 °C) |
Operation | Foil keypad |
Display | backlit LCD (on keypress) |
Suitable Vials | Round Cuvettes 24 mm |
Calibration | factory calibration and user calibration. Reset to factory calibration possible |
Interfaces | Infrared |
Internal Storage | internal ring memory for 16 data sets |
Power Supply | 4 micro batteries (AAA) |
Battery Life Time | approx. 17 hours |
Auto – OFF | Yes |
Clock | Real Time Clock and Date |
Portability | Portable |
Environmental Conditions | Temperature: 5 - 40 °C rel. humidity: 30 - 90 % (non condensing) |
Compliance | CE |
Protection Class | IP 68 |
Dimensions | 155 x 75 x 35 mm |
Weight | 260 g |
Comparison between Lovibond portable photometers MD 100, MD 110 and MD 200 can be seen below.
All these photometers measurements using high-quality interference filters with long-life LEDs as a light source in a transparent sample chamber. The units provide accurate, reproducible results very quickly. Other major advantages include ease of operation, ergonomic design, compact dimensions and safe handling. Using an internal ring memory, the last 16 data sets (MD100, MD200) and 125 data sets (MD110) are stored automatically with date, time, parameter and measurement value. The tests are conducted using either Lovibond® tablet reagents with long-term stability, VARIO powder reagents or liquid reagents.
The benchtop photometer MD 100 from Lovibond combines high-quality interference filters with long-term stable LEDs as light sources in a high-precision optic. This provides precise measurement results at all times - no matter where you use it. Below are all the potential combinations for MD 200, please enquire for products not available online:
Parameter Count | Title | Parameter and Range | Part Number |
---|---|---|---|
Single | MD 100 Aluminium, tablet reagents | Aluminium T 0.01 - 0.3 mg/l Al | 276200 |
Single | MD 100 Aluminium, powder reagents | Aluminium PP 0.01 - 0.25 mg/l Al | 276205 |
Single | MD 100 Ammonia, tablet reagents | Ammonia T 0.02 - 1 mg/l N | 276060 |
Single | MD 100 Ammonia, powder reagents | Ammonia PP 0.01 - 0.8 mg/l N | 276065 |
Single | MD 100 Chlorine, tablet reagents | Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) |
276000 |
Single | MD 100 Chlorine, liquid reagents | Chlorine L 0.02 - 4.0 mg/l Cl2 a) | 276005 |
Single | MD 100 Chlorine DUO, tablet reagents | Chlorine HR T 0.1 - 10 mg/l Cl2 a) Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) |
276020 |
Single | MD 100 Chlorine DUO, powder reagents | Chlorine HR PP 0.1 - 8 mg/l Cl2 a) Chlorine PP 0.02 - 2 mg/l Cl2 a) |
276025 |
Single | MD 100 Chlorine, powder reagents | Chlorine PP 0.02 - 2 mg/l Cl2 a) | 276010 |
Single | MD 100 Chlorine HR (potassium iodide), tablet reagents | Chlorine HR (KI) T (105) 5 - 200 mg/l Cl2 | 276170 |
Single | MD 100 Chlorine dioxide, tablet reagents | Chlorine dioxide T 0.02 - 11 mg/l ClO2 | 276030 |
Single | MD 100 Chlorine dioxide, powder reagents | Chlorine dioxide PP 0.04 - 3.8 mg/l ClO2 | 276035 |
Single | MD 100 Chloride, tablet reagents | Chloride T 0.5 - 25 mg/l Cl- Chloride T 5 - 250 mg/l Cl- i) |
276180 |
Single | MD 100 COD, tube tests | COD HR TT 200 - 15000 mg/l CODb) COD LMR TT 15 - 300 mg/l CODb) COD LR TT 3 - 150 mg/l CODb) COD MR TT 20 - 1500 mg/l CODb) |
276120 |
Single | MD 100 Iron, tablet reagents | Iron T 0.02 - 1 mg/l Fe | 276050 |
Single | MD 100 Iron TPTZ, powder reagents | Iron (TPTZ) PP 0.02 - 1.8 mg/l Fe | 276055 |
Single | MD 100 Iron, powder reagents | Iron PP 0.02 - 3 mg/l Feg) | 276056 |
Single | MD 100 Fluoride (without reagents) | Fluoride L 0.05 - 2 mg/l F- | 276090 |
Single | MD 100 Hardness, total, tablet reagents | Hardness total HR T 20 - 500 mg/l CaCO3 i) Hardness total T 2 - 50 mg/l CaCO3 |
276190 |
Single | MD 100 Urea, tablet reagents | Urea T 0.1 - 2.5 mg/l Urea Urea T 0.2 - 5 mg/l Ureai) |
276210 |
Single | MD 100 Hazen/A.P.H.A., no reagents required | Hazen 24 10 - 500 mg/l Pt | 276160 |
Single | MD 100 Copper, tablet reagents | Copper T 0.05 - 5 mg/l Cua) | 276080 |
Single | MD 100 Copper, powder reagents | Copper PP 0.05 - 5 mg/l Cu | 276085 |
Single | MD 100 Manganese LR, tablet reagents | Manganese T 0.2 - 4 mg/l Mn | 276100 |
Single | MD 100 Manganese LR, powder reagents | Manganese LR PP 0.01 - 0.7 mg/l Mn | 276105 |
Single | MD 100 Manganese HR, powder reagents | Manganese HR PP 0.1 - 18 mg/l Mn | 276106 |
Single | MD 100 Molybdenum LR, powder reagents/ tablet reagents | Molybdate LR PP 0.05 - 5 mg/l MoO4 | 276140 |
Single | MD 100 Molybdenum HR, powder reagents | Molybdate HR PP 0.3 - 40 mg/l MoO4 | 276141 |
Single | MD 100 Molybdenum, tablet reagents | Molybdate T 0.6 - 50 mg/l MoO4 | 276142 |
Single | MD 100 Phosphate, tablet reagents | Phosphate LR T 0.02 - 1.3 mg/l P | 276040 |
Single | MD 100 Phosphate, powder reagents | Phosphate PP 0.02 - 0.8 mg/l P | 276045 |
Single | MD 100 Silica, tablet reagents | Silcate T 0.05 - 4 mg/l SiO2 | 276110 |
Single | MD 100 Silica LR, powder reagents | Silicate LR PP 0.1 - 1.6 mg/l SiO2 | 276115 |
Single | MD 100 Silica HR, powder reagents | Silicate HR PP 1 - 90 mg/l SiO2 | 276116 |
Single | MD 100 Suspended solids, no reagents required | Suspended solids 24 10 - 750 mg/l TSS | 276150 |
2in1 | MD 100 Chlorine, pH, tablet reagents (OTZ) | Chlorine HR T 0.1 - 10 mg/l Cl2 a) Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) pH value L 6.5 - 8.4 pH-value T 6.5 - 8.4 |
278020 |
2in1 | MD 100 Chlorine, pH, liquid reagents (OTZ) | Chlorine L 0.02 - 4.0 mg/l Cl2 a) pH value L 6.5 - 8.4 |
278025 |
2in1 | MD 100 Chlorine, pH, powder reagents for chlorine | Chlorine HR PP 0.1 - 8 mg/l Cl2 a) Chlorine PP 0.02 - 2 mg/l Cl2 a) pH value L 6.5 - 8.4 pH-value T 6.5 - 8.4 |
278030 |
3in1 | MD 100 Chlorine, pH, Cyanuric acid, tablet reagents (OTZ) | Chlorine HR T 0.1 - 10 mg/l Cl2 a) Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) CyA T 10 - 160 mg/l CyA pH value L 6.5 - 8.4 pH-value T 6.5 - 8.4 |
278010 |
3in1 | MD 100 Chlorine, pH, Cyanuric acid, liquid reagents for Chlorine and pH (OTZ) | Chlorine L 0.02 - 4.0 mg/l Cl2 a) CyA T 10 - 160 mg/l CyA pH value L 6.5 - 8.4 |
278015 |
3in1 | MD 100 Chlorine, pH, Alkalinity-m, tablet reagents (OTZ) | Alkalinity-m T 5 - 200 mg/l CaCO3 Chlorine HR T 0.1 - 10 mg/l Cl2 a) Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) pH value L 6.5 - 8.4 pH-value T 6.5 - 8.4 |
278060 |
3in1 | MD 100 Chlorine, pH, Alkalinity-m, liquid reagents for Chlorine and pH (OTZ) | Alkalinity-m T 5 - 200 mg/l CaCO3 Chlorine L 0.02 - 4.0 mg/l Cl2 a) pH value L 6.5 - 8.4 |
278065 |
2in1 | MD 100 Chlorine LR, Chlorine HR, Chlorine dioxide, tablet reagents | Chlorine dioxide T 0.02 - 11 mg/l ClO2 Chlorine HR T 0.1 - 10 mg/l Cl2 a) Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) |
278000 |
4in1 | MD 100 Chlorine, pH, Cyanuric acid, Alkalinity-m, tablet reagents (OTZ) | Alkalinity-m T 5 - 200 mg/l CaCO3 Chlorine HR T 0.1 - 10 mg/l Cl2 a) Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) CyA T 10 - 160 mg/l CyA pH value L 6.5 - 8.4 pH-value T 6.5 - 8.4 |
278070 |
4in1 | MD 100 Chlorine, pH, Cyanuric acid, Alkalinity-m, liquid reagents for Chlorine and pH (OTZ) | Alkalinity-m T 5 - 200 mg/l CaCO3 Chlorine L 0.02 - 4.0 mg/l Cl2 a) CyA T 10 - 160 mg/l CyA pH value L 6.5 - 8.4 |
278075 |
4in1 | MD 100 Chlorine DUO, Chlorine HR, pH, Alkalinity-m, Calcium hardness (OTZ) | Alkalinity-m T 5 - 200 mg/l CaCO3 Chlorine HR (KI) T (105) 5 - 200 mg/l Cl2 Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine MR PP 0.02 - 3.5 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) Hardness Calcium (B) T 20 - 500 mg/l CaCO3 pH value L 6.5 - 8.4 pH-value T 6.5 - 8.4 |
278160 |
5in1 | MD 100 Chlorine, pH, Cyanuric acid, Alkalinity-m, Calcium hardness, tablet reagents (OTZ) | Alkalinity-m T 5 - 200 mg/l CaCO3 Chlorine HR T 0.1 - 10 mg/l Cl2 a) Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) CyA T 10 - 160 mg/l CyA Hardness Calcium (B) T 20 - 500 mg/l CaCO3 pH value L 6.5 - 8.4 pH-value T 6.5 - 8.4 |
278080 |
6in1 | MD 100 Chlorine, Bromine, pH, Cyanuric acid, Alkalinity-m, Calcium hardness, tablet reagents (OTZ) | Alkalinity-m T 5 - 200 mg/l CaCO3 Bromine T 0.05 - 13 mg/l Br2 Chlorine HR T 0.1 - 10 mg/l Cl2 a) Chlorine L 0.02 - 4.0 mg/l Cl2 a) Chlorine T 0.01 - 6.0 mg/l Cl2 a) CyA T 10 - 160 mg/l CyA Hardness Calcium (B) T 20 - 500 mg/l CaCO3 pH value L 6.5 - 8.4 pH-value T 6.5 - 8.4 |
278090 |
Boiler Water | MD 100 Boiler Water | Aluminium PP 0.01 - 0.25 mg/l Al Chloride L (A) 0.5 - 20 mg/l Cl- Copper T 0.05 - 5 mg/l Cua) DEHA PP 0.02 - 0.5 mg/l DEHA Hydrazine P 0.05 - 0.5 mg/l N2H4 Iron LR L (A) 0.03 - 2 mg/l Fe Oxygen dissolved C 10 - 800 µg/l O2 c) Phosphate HR L 5 - 80 mg/l P Polyacrylate L 1 - 30 mg/l Polyacryl Silicate HR PP 1 - 90 mg/l SiO2 |
276230 |
Cooling Water | MD 100 Cooling Water | 16 Parameters | 276240 |
Have you considered your servicing requirements? Much like a car, scientific instruments may need a bit of care to keep them working reliably and at their most accurate.
The Camlab Service Department have over 60 years experience in scientific instrumentation. We have a number of highly qualified internal and field-based service engineers able to perform a variety of jobs including preventative maintenance servicing and calibration.
Camlab can offer yearly service contracts which include two main types of preventative maintenance/calibration services – the back to base service and an engineer visit to your site. These options are can be offered with differing turn around times depending on the instrument type and your requirements along with loan instruments if available.
|
|
A typical preventative maintenance service includes the following: Inspection and testing of the unit
Dismantling, cleaning and inspection of the key parts
Correction of minor defects including adjustments if necessary
Reassembly and testing
Take post adjustment / cleaning readings
Supply of calibration certificate
Return of the item to you (back to base services)
|
Request a quote for your servicing requirements today.
Products in this Range