The Gastec smoke test kit allows the user to monitor airflow movements within the work place.
Break off both ends of the smoke tube and attach a rubber squeeze bulb to one end of the tube. When the bulb is squeezed, a chemical reaction produces white smoke. This smoke makes the air flow (in ventilation systems/ducts, etc.) visible
The Gastec smoke test kit is ideal for indoor air measurement. Both ends of the smoke tube are broken and connected to a rubber bulb. Moisture in the air reacts with the reagent to produce a white smoke. A single tube can be used to perform in excess of 50 tests by sealing the tube ends with the rubber caps provided.
Ideal to validate fume hoods or extraction systems, monitor ventilation systems or air ducts.
GASTEC gas detector tubes are thin glass tubes with calibration scales printed on them so you can directly read the concentrations of the substances (gases and vapours) to be measured. Each tube contains a particulate matrix (e.g. silica gel, alumina), which binds carefully selected and highly stable detection reagents that are especially sensitive to the target substance to produce a distinct layer of colour change. The tubes are hermetically sealed at both ends.
Currently, over 600 types of target gas can be measured.
Most GASTEC tubes can be used in the temperature range of 0 - 40ºC and relative humidity range of 0- 90%. However, a few tubes have different temperature or humidity ranges. Please refer to the instruction manuals before use.
The handy design GASTEC GV-100 cylinder pump, with an inner capacity of 100mL, weighs only 240g. It is designed for use with almost all GASTEC short-term measurement detector tubes.
The following table provides an overview and comparison of JIS K 0814 and major international industrial standards.
Standard / Requirements |
JIS K 0804:2022 | ISO17621:2015 | lUPAC (published in 1982) |
ANSl / ISEA 102-1990 (R2015) |
||
---|---|---|---|---|---|---|
Sampling pump | Volume | ≦±5% | ≦±5mL | ≦±5% | ≦±5% | |
Air leakage | ≦3% | ≦3mL/min | ≦ 3%/min | ≦ 3%/min | ||
Durability | 1000 pump strokes | 1000 pump strokes | - | 100 pump strokes | ||
Detector tube | Obliqueness of the boundary discolouration | ≦20% | ≦20% | ≦20% | ≦20% | |
Measurement performance |
Class A B* ≦15% (Conc. of 33.3% or more of MS***) ≦15 to 25%(Conc. of 20 to 33.3% of MS***) ≦25% (Conc. of 20% or less of MS***) (B*+U**) ≦30% (Conc. of 26.7% or more of MS***) ≦30 to 50% (Conc. of 16 to 26.7% of MS***) ≦50% (Conc. of 16 % or less of MS***) Class B (B*+U**)≦50%
|
U**≦50% |
Class A Read value≦± 25% (1, 2 and 5×MPC****) ≦± 35% (0.5×MPC****)
Read value ≦± 35% (0.5, 1, 2 and 5×MPC****) |
Read value ≦± 25% (1, 2 and 5×MPC****) ≦± 35% (0.5×MPC****) |
*B: Bias of the scale reading against the test gas concentration
**U: Expanded uncertainty for a detector tube measurement system (The calculations of Expanded uncertainty are different for each standard.)
***MS: Maximum scale value of the detector tube
****MPC: Maximum permissible concentration, e.g. TLVs (Threshold limit values), etc.
Related Products
Check items to add to the cart